Preference and speech recognition thresholds of self-adjusted hearing aids using a model-based parameter space
* Presenting author
Abstract:
Self-adjustment is one approach to stronger involve the user in the process of hearing aid fitting. To define an individual parameter space for a specific user, auditory models can be employed to only include settings that ensure a minimum speech intelligibility and maximum loudness. A self-adjustment experiment was conducted with 10 experienced hearing aid users with moderate to severe hearing impairments. After self-adjusting the hearing aid setting, aided speech recognition thresholds (SRTs) were measured with the German Matrix Test. Additionally, subjects had to compare their chosen settings to a setting derived by a standard prescription rule in a combined discrimination and classification task. This study aims at verifying the premise that all chosen settings lead to SRTs lower than -3 dB SNR. It was also hypothesized that self-adjusted settings of hearing aids would be preferred over standard prescribed settings. Results showed that 85% of the measured SRTs were below -3 dB SNR, indicating that the model-based approach to ensure a certain speech intelligibility is generally feasible. However, subjects tended to prefer the prescribed hearing aid setting over the self-adjusted one. This might be caused by their substantial experience with hearing aids using prescribed fitting.