Vibration Characteristics of Gradient Coil in Low Field MRI Scanners

* Presenting author
Day / Time: 18.03.2025, 15:20-16:00
Typ: Poster
Information: The posters will be exhibited in Hall E north from Tuesday to Thursday, sorted by thematic context in the poster island indicated in the session title. The poster session at the specified time offers the opportunity to enter into discussion with the authors.
Abstract: Magnetic resonance imaging (MRI) is an important tool for diagnosing the human body and many biological as well as non-biological materials. The MRI scans are advantageous because it does not involve use of ionizing radiations. The vibration associated with the gradient coil of MRI systems during image acquisition may result in compromised image quality and this problem is more severe for low field MRI with lower scanning resolution. This study investigates the vibration characteristics of the lower gradient coil set of a low field open configuration MRI system, using finite element analysis and experimental measurement. The important frequencies of vibration are identified which is helpful to develop vibration mitigation strategy.